Balance Configuration Setup
Compound Direct supports up to 40 balance brands, with many models tested and verified by our team to ensure reliable communication during compounding workflows. While the balances listed in this guide have been physically validated, other compatible models may also work depending on their communication capabilities.
This guide provides configuration instructions for each supported balance model, including the communication settings required to connect successfully with Compound Direct. Whether you are setting up a new balance, reconfiguring an existing device, or troubleshooting a connection issue, these instructions will help ensure your balance is configured correctly for reliable weight transmission.
With this guide, you can:
- Identify whether your balance is a tested and supported model.
- Configure the required communication settings for your balance.
- Connect your balance to Compound Direct for reliable real-time weight capture.
Once you've identified your balance model, follow the setup instructions below to configure it for use with Compound Direct.
Supported balances
If your balance is not listed here it is likely still supported as we support up to 40 brands of balances, however these are a list of balances that have been physically tested and verified by our team.
- • A&D FX Series
- • A&D FZ Series
- • A&D GX Series
- • A&D GF Series
- • A&D EK Series
- • A&D HR Series
A&D (GX/GF/FZ/FX/HR)
Use the A&D balance simulator to understand how to configure your balance.
-
Ensure the balance is turned on.
-
Hold down
sampleto enter menu. Once entered you should see bA5Fnc. -
Press
sample4-5 times to navigate to dout option. -
Press
printto enter dout submenu. You should see Prt (0) as the key mode. -
Press
re-zerothree times to set to the streaming mode to (3). -
Press
print, this will show END and then after one second show SiF. -
Press
printto enter SiF submenu. You should see bPS (2).-
For some models (A&D GX6038), this may go to mode first and you need to press sample to go to bPS.
-
-
Press
re-zerotwice to set baud (bPS) to 9600 (4). -
Press
samplego to the next option. You should see btPr (0). -
Press
re-zerotwice to set to 8bits and none (2). -
Press
samplego to the next option. You should see CrLF (0). -
Press
samplego to the next option. You should see tYPE (0).-
If type is 1, press re-zero until it reaches 0.
-
-
Press
printto save. -
Press
calibrateto exit.
A&D (EK)
- Turn OFF the balance.
- With the power turned OFF, press and hold the
re-zerokey and press theon/offkey to turn the power ON and enter the function setting mode. If done correctly, you should see Func. - Press
printto enter Func submenu. You should see PoFF 0. - Press
sample6 times to navigate toPrtoption. - Press
re-zerotwo times to set to streaming (4). - Press
sample2-3 times to navigate to BPS (0). - Press
re-zerotwice to set to 9600 (2). - Press
samplego to the next option. You should see btPr (0). - Press
re-zerotwice to set to 8bits and none (2). - Press
printto save. This will show End and then CP Hi. - Press
offandonto restart the balance.
Mettler Toledo (AB104-S/PB4)
- In weighing mode, press and hold down the
«Cal/Menu»key until “MENU” appears in the display. Release the key. - Press
«Cal/Menu»until you reach 'PrintEr' or until your reach "HOST". - Press
F repeatonce until it says 'HOST'. - Press
«Cal/Menu»twice until you reach 'S-oFF' or 'S. Stb' or 'S. Cont'. - Press
F repeatuntil it says 'S. Cont'.-
It can be either set to 'Continuos' or 'Stable'. If you pref the weights to be displayed only when Stable, select the stable option.
-
- Press
«Cal/Menu»once until you reach 'bd 2400' or bd XXXX. - Press
F repeatuntil it says 'bd 9600'. - Press
«Cal/Menu»once until you reach '7b-E' or '7b-no' or '8b-no' or '7b-odd'. - Press
F repeatuntil it says '8b-no' - Hold down
«Cal/Menu»until it shows 'Stored' on the display and then release. - If the balance does not appear immediately, retry the steps above and ensure the following settings are correct: HOST, S.Cont, bd 9600, 8b-no.
Mettler Toledo (ML)
- In weighing mode, press and hold down the
settingskey until it shows BASIC on the display. Release the key. - Use the navigation to go to Interface.
- Navigate to RS232C. You should see HOST (default), if you see Printer, change it to HOST.
- Navigate to HOST.
- You should see Std.Stb (default). Change it to Std.Cont.
- Press
(C)backto exit. - You will be prompted to save settings:
- Change Save: NO to Save: YES by pressing the
repeatkey. - Press
Returnto confirm.
- Change Save: NO to Save: YES by pressing the
- Ensure the following default settings are correct:
- Baud rate: 9600
- Data bits/parity: 8b-no
Mettler Toledo (XPR)
Navigation: Balance menu → Settings → LabX / Services
- Navigate to the LabX / Services section.
- Set MT-SICS to USB.
- Once done, tap
Save.
OHaus Pioneer/Ranger
- Hold down
menuuntil it says Cal. - Press
No/Printuntil Print is displayed. - Press
Yes/Zeroto enter Print Menu. - Press
Nountil we get to the AutoPrint/APrint option. - Press
Yesto enter AutoPrint. - Press
Nountil the Continuous option. - Press
Yesto confirm. - Restart the balance.
OHaus Defender
- Open Menu (use the arrow_drop_up and arrow_drop_down buttons and
tickto move forward into sub menu). - Press
downto 7 Communication, presstick. - Press
downonce to Print Setup, presstick. - Press
tickfor assignment. - Press
tickdemand option, cycle through options, by pressing down to OH Continous. - Press
exitbutton (circle with arrow out). - Restart the balance.
OHaus Adventurer
- Open communication settings.
- Tap
print settings. - Set auto print continuous.
- Restart the balance.
Precisa (XT/XB)
- Press
«ON/OFF»to switch on the balance.- Continue to hold down the
MENU / ESCAPEbutton during the start-up.
- Continue to hold down the
- Press
arrow_drop_downuntil you reach 'SET DATA PRINT'.- Press
arrow_rightto select this item. - Menu should say "AUTO-START".
- Press
- Press
arrow_drop_downuntil you reach 'SET PRINTFORMAT'.- Press
arrow_rightto select this item. - Menu should say "DATE AND TIME".
- Press
- Press
arrow_drop_downuntil you reach 'GROSS AND TARE'.- Press
arrow_rightto select this item. - Menu should say "OFF".
- Press
- Press
arrow_drop_downuntil you reach 'ON'. - Press
arrow_leftto return. - Press
arrow_drop_downuntil you reach 'SET INTERFACE'.- Press
arrow_rightto select this item. - Menu should say "BAUDRATE".
- Press
- Press
arrow_rightto select this item.- Menu should say "600".
- Press
arrow_drop_downuntil you reach '9600'. - Press
arrow_leftto return.- Menu should say "PARITY".
- Press
arrow_rightto select this item.- Menu should say "7-EVEN-1STOP".
- Press
arrow_drop_downuntil you reach '8-NO-1STOP'. - Press
arrow_leftto return. - Press
Menu / ESCAPEto leave the menu.
Precisa (ES)
- Press
«ON/OFF»to switch on the balance.- Continue to hold down the
MENU / ESCAPEbutton during the start-up. - Process (approximately 10 seconds), until “UNIT-1” appears in the display.
- Continue to hold down the
- Press
arrow_drop_downuntil you reach 'SET DATA PRINT'. - Press
arrow_rightto select this item.- Menu should say "AUTO-START".
- Press
arrow_rightto select this item.- Menu should be flashing 'OFF'.
- Press
arrow_drop_downto change 'OFF' to 'ON'. - Press
tarebutton to confirm. - Press
arrow_leftto return. - Press
arrow_drop_downto 'MODE'.- Menu should say "STABLE".
- Press
arrow_rightto select this item. - Press
arrow_drop_downtwice to change to 'CONTINUOUS'. - Press
arrow_leftto return. - Press
MENU / ESCAPEto leave the menu.
Radwag X2 (AS/PS/WLC)
- Ensure the balance is turned on.
- Press the
Personbutton (4th icon from right at the bottom):- Select Admin.
- Enter default password
1111.
-
If you have setup a custom admin password enter that instead. If the password is unknown you may need to factory reset the balance.
- Press the
Tickin the bottom right to confirm.- Ensure the balance user is "Admin" in the bottom right corner under "User:".
- Press the
settingsbutton (bottom left corner). - Scroll down slightly and press the
Communicationbutton.- Ensure COM1 is selected.
- Press
back(middle left). - Scroll down slightly and press the
Peripheralsbutton. - Press the
Computerbutton.- Select 'Continuous Transmission'.
- Select 'Current Unit'.
- Press
back. (middle left). - Press the
Printerbutton.- Set the Baud Rate to 9600.
- Ensure the Data Bits is "8", Stop bits is "1" and Parity is "None".
- Press
back(middle left). - Press
backagain to return to normal weighing. - Ensure the cable provided is connected to the COM1 Port.
Sartorius Quintix
No configuration needed
Sartorius Entris
No configuration needed
Sartorius Cubis
No configuration needed
Shimadzu
Reset balance
- Press
Caluntil you reach 'SELEAUS'. - Press
Caluntil 'S' is flashing. - Press
O/Tto select. - Press
Caluntil you reach 'r'. - Press
O/Tto select - "Reset?" will be displayed. - Press
O/Tto select. - Press power until you read
backto weighing mode.
Turn on auto-print
- Press
Caluntil you reach 'SELEAUS'. - Press
Caluntil 'A' is flashing. - Press
O/Tto select. - Press
Caluntil until 'ap' is flashing. - Press
O/Tto select. - Press
Cal5 timtes till it says 'on Cont'. - Press
O/Tto select "SET" will be displayed. - Press
poweruntil you readbackto weighing mode.
Chance communication speed to 9600 bits per second
- Press
Calmenu. - Press
Caluntil you reach 'SELEAUS'. - Press
Caluntil the top right lightning/communication symbol is flashing. - Press
O/Tto select, -HFbpSd will be shown. - Press
Caluntil you reach 'b'. - Press
O/Tto select. - Press
Caluntil you reach b- 9600. - Press
O/Tto select, "SET" will be displayed. - Press
poweruntil you read back to weighing mode.
Activate
- PRESS
PRINTto activate. - Select the
menukey at the top left of the display. - Press the
downbutton until you reach the DEVICE parameter - Press the
[Confirm]button. (far bottom right button)- You should see DAT.REC.
- Press the
downbutton until you reach the PARITY parameter.- You will see ODD as default.
- Press the
[Confirm]button. (far bottom right button).- Change to NONE.
- Exit the menu by pressing
back(far bottom left button).
Shimadzu (UW/UX/BX/TX)
- Press the
[MENU]key twice from the gram-dis-play.- “Setwin” appears.
- Press the
[O/T]key. Verify the stability mark is illuminated with “Setwin” display.- All the communication settings for WindowsDirect have been made.
- Go to “STAND-BY” by pressing the
[POWER]key several times and unplug the AC adapter from the balance.- Unpluging the balance once is necessary after the above setting.
- Connect the RS-232C cable to the balance.
- Turn on the balance.
Vibra (AB/AJ)
-
Ensure the balance is turned on.
-
Press and hold down the
Functionkey until the indicator changes to “FUNC” then release the key.- Once complete, you should see 1 b.G.
-
Press
Functionsix times to navigate to 7 1.F. (1) option. -
Press the
tarebutton three times, until it reaches 7 1.F. (4).-
For AJ Models only press the tare button once, until it reaches (2).
-
-
Press function once, it should display 7 1 o.c. (7) option.
-
Press the
tarebutton two times, until it reaches 7 1 o.c. (1). -
Press
Functiononce, it should display 7 2 b.L. (1) option. -
Press the
tarebutton three times, until it reaches 7 2 b.L. (4). -
Press
tareto close the menu .
Vibra (ALE)
- Press
downuntil we reach external i/o. - Press
rightto get to rs232. - Press
rightto get to ACTIVATE.- It should say OFF, press
f4 (enter)to set to ON.
- It should say OFF, press
- Press
downto get to FORMAT.- It should say CBM.
- Press
downto get to CONDITION.- It should say 7 press
f4 (enter)util it reaches '2' (print on stable).
- It should say 7 press
- Press
downtwice until we reach BAUD RATE.- It should say 1200, press
f4 (enter)until we reach 9600.
- It should say 1200, press
- Press
downto PARITY. - Press
downto STOP BIT.- It should say 2BIT, press
f4 (enter)to set to 1BIT.
- It should say 2BIT, press